Where do I send the manual for Alcon Constellation vision system?

Portable and mobile RF communications equipment can affect this medical electrical equipment. 8065751025 ix fConstellation Vision System Comments or corrections concerning this manual should be addressed to: Alcon Technical Services Group PO BOX 19587 Irvine, CA, USA 92623 All rights reserved.

What are the guidelines for use of constellation® vision system?

? Good clinical practice dictates the testing for adequate irrigation, aspiration ?ow, and operation as applicable for each handpiece prior to entering the eye. ? Do not use the Constellation Vision System system near ?ammable anesthetics. ? Use only Alcon-supplied A.C. power cords.

What is the source air pressure for Constellation vision system?

Source Air Pressure Requirements The Constellation Vision System is designed to operate using different levels of source air pressure. The system operates automatically with pressures of 58 (4 bar) to 120 psi (8.3 bar). Between 4 bar and 5 bar, vacuum performance is reduced.

What is the function of the Illuminator on a constellation?

Illuminator Function Housed inside Constellation tabletop, the illuminator provides two channels of illumination from a single xenon arc lamp. Two front panel sockets accept ACMI compatible fiber optic light guides to provide intraocular illumination.